Particles contained reduced sulfur can be detected on environmental tapelifts
by placing a droplet of sodium azide/iodine solution on the
particle side of the tape and observing the tape under the stereomicroscope.
Even micrometer sized particles will generate visible bubbles of
nitrogen in a few minutes. This is primarily for industrial sites because
skin flakes and hair contain reactable sulfur in their proteins.
|